Abstract

Urine drug screen (UDS) immunoassays are a quick and inexpensive method for determining the presence of drugs of abuse. Many cross-reactivities exist with other analytes, potentially causing a false-positive result in an initial drug screen. Knowledge of these potential interferents is important in determining a course of action for patient care. We present an inclusive review of analytes causing false-positive interferences with drugs-of-abuse UDS immunoassays, which covers the literature from the year 2000 to present. English language articles were searched via the SciFinder platform with the strings ‘false positive [drug] urine’ yielding 173 articles. These articles were then carefully analyzed and condensed to 62 that included data on causes of false-positive results. The discussion is separated into six sections by drug class with a corresponding table of cross-reacting compounds for quick reference. False-positive results were described for amphetamines, opiates, benzodiazepines, cannabinoids, tricyclic antidepressants, phencyclidine, lysergic acid diethylamide and barbiturates. These false-positive results support the generally accepted practice that immunoassay positive results are considered presumptive until confirmed by a second independent chemical technique.

Introduction

Immunoassays dominate urine drug screens (UDSs) because they are simple to use, easy to automate and provide rapid results (1). Unfortunately, they are subject to cross-reactivity with structurally related and unrelated compounds potentially yielding false-positive results. Further complicating the issue are the many available platforms with differing cross-reactivities. Immunoassays for selected drug classes, e.g., opiates and benzodiazepines, are also subject to clinically important false negatives (2). False-negative results can be caused by a variety of factors including the cross-reactivity of the antibody used by the assay, the cutoff concentration for a positive result and length of time between drug ingestion and specimen acquisition. False negatives are not covered in this review but present opportunities for significant patient mismanagement if not understood. The best practice following a positive UDS involves confirmation with the mass spectrometry (MS) technique such as gas chromatography–mass spectrometry (GC-MS) or liquid chromatography–tandem mass spectrometry. Regrettably, MS testing is limited or nonexistent in many hospital laboratories. When confirmatory testing is performed, results are generally unavailable for several days. Owing to the delay in receiving confirmation results, decisions about patient care are frequently made on ‘presumptive positive’ drug screening results.

Because many providers have limited knowledge of immunoassay cross-reactivity data, patients with false-positive results may lose eligibility in rehabilitation programs, be inappropriately terminated from employment or suffer from medical staff bias because of lack of trust (3). Although this topic has been reviewed previously (24), our aim is to provide a concise, comprehensive and up-to-date account of substances potentially interfering with UDS immunoassays. This review is meant to serve as a guide for practitioners to assess potential false-positive UDS results while waiting for confirmatory test results to become available.

This review focuses on cases where the cause of a false positive is identified. In practice, there are many cases where immunoassay positive results are not confirmed when analyzed with a more sensitive and specific technique. While it is useful to understand some of the common causes of false-positive immunoassay results, in the majority of cases the cause of false-positive screening results is unknown. Owing to the number of false positives, all immunoassay results are considered ‘presumptive’ until confirmed by an independent chemical technique.

Materials and methods

An inclusive literature review was conducted for five predefined drugs of abuse classes and a miscellaneous drug class that included several different drugs. The predefined drug classes include amphetamines, tricyclic antidepressants (TCAs), benzodiazepines, cannabinoids and opiates (natural and synthetic). The drug classes were then investigated using the SciFinder platform that includes MEDLINE and CAPLUS databases. The strings ‘false positive [drug] urine’ for English language articles were searched. In addition, a panel of over 60 common over the counter medications, prescription drugs and their metabolites, and illicit drugs were included using the same search syntax. To maintain current relevance, only articles published after the year 2000 were included in the search. This search yielded 173 articles. These articles were then analyzed to determine if they included data on false-positive urine immunoassay results. A total of 62 articles met these criteria and were included in this review. Any articles prior to 2000 included in this review are used to clarify current cross-reactivity issues.

Publications describing interferences with UDS present differing degrees of evidence to support their claims. The weakest evidence is noted by simply stating that the patient(s) had exposure to the proposed interfering drug, or comparing two-dimensional drug structures for similarity or dissimilarity between the proposed interferent and the assay target drug/drug class. Stronger evidence is obtained by adding a pure standard of the suspected interferent drug (and/or potential metabolites) at several concentrations to drug free urine (DFU) and testing the immunoassay(s) in question. Weaker variants of this protocol include adding crushed tablets or capsules instead of pure drug standard and using water or organic solvent/water mixtures instead of human DFU as a matrix. Useful supportive evidence may include quantitation of potentially interfering substances in patient specimens and chart reviews, but these methods are not a substitute for directly testing the proposed interferent in the appropriate matrix. Some studies dose subjects with the drug in question and evaluate immunoassays before and after dosing. In these cases, positive results are analyzed by confirmatory testing methods to demonstrate the absence of the target compound. The articles presented here for drugs causing false-positive immunoassay results include all of the above methods.

Amphetamines

The abuse of amphetamines largely stems from their intense euphoria-inducing effects (5). Because amphetamines have significant toxicity and are widely abused, they are commonly included in routine UDSs. Amphetamine and methamphetamine (1), as shown in Figure 1, are simple molecules that make it difficult to develop antibodies that are specific to these drugs. In addition, there are many structurally related sympathomimetics that are commonly ingested. For these reasons, immunoassays designed to detect amphetamines have historically been subject to significant cross-reactivity problems. Table I lists compounds that have been reported to cause false-positive results with currently used immunoassays.
Table I

False-Positive Results for Amphetamine Immunoassays

Cross-reacting drugImmunoassay platform (positive cutoff)Immunoassay nameLevel which false (+) occurredLevel of evidenceReference
BupropionSiemens Dimension®SYVA EMIT II Plus®OccurredRetrospective chart review(6)
ChlorpromazineRoche Hitachi 911® (1.0 μg/mL cutoff)SYVA EMIT-MAM®OccurredRetrospective chart review(1)
DMAARoche Modular P® (300 ng/mL cutoff)Roche kinetic interaction of microparticles in a solution
(KIMS®)
7.5 μg/mLSpiking of standards in DFU(7)
SYVA EMIT II Plus®3.1 μg/mLSpiking of standards in DFU(7)
Roche kinetic interaction of microparticles in a solution
(KIMS)® and SYVA EMIT II®
6.9 μg/mLTherapeutic dosing(7)
LabetalolNot mentionedNot mentionedOccurredCase report(8)
MetforminBiosite Triage®OccurredCase report(9)
OfloxacinTdxFlx® (300 ng/mL cutoff)AM/MA II®OccurredTherapeutic dosing(10)
PromethazineRoche Hitachi 911® (1.0 μg/mL cutoff)SYVA EMIT-MAM®OccurredRetrospective chart review(1)
Trazodone (m-CPP)Olympus UA 400®SYVA Ecstasy
EMIT II®
3 μg/mL
(m-CPP)
Spiking of standards in DFU(11)
Roche Cobas c501®, cutoff 300 μg/LAmphetamines
II®
6.7 μg/mL
(m-CPP)
Native matrix(12)
Cross-reacting drugImmunoassay platform (positive cutoff)Immunoassay nameLevel which false (+) occurredLevel of evidenceReference
BupropionSiemens Dimension®SYVA EMIT II Plus®OccurredRetrospective chart review(6)
ChlorpromazineRoche Hitachi 911® (1.0 μg/mL cutoff)SYVA EMIT-MAM®OccurredRetrospective chart review(1)
DMAARoche Modular P® (300 ng/mL cutoff)Roche kinetic interaction of microparticles in a solution
(KIMS®)
7.5 μg/mLSpiking of standards in DFU(7)
SYVA EMIT II Plus®3.1 μg/mLSpiking of standards in DFU(7)
Roche kinetic interaction of microparticles in a solution
(KIMS)® and SYVA EMIT II®
6.9 μg/mLTherapeutic dosing(7)
LabetalolNot mentionedNot mentionedOccurredCase report(8)
MetforminBiosite Triage®OccurredCase report(9)
OfloxacinTdxFlx® (300 ng/mL cutoff)AM/MA II®OccurredTherapeutic dosing(10)
PromethazineRoche Hitachi 911® (1.0 μg/mL cutoff)SYVA EMIT-MAM®OccurredRetrospective chart review(1)
Trazodone (m-CPP)Olympus UA 400®SYVA Ecstasy
EMIT II®
3 μg/mL
(m-CPP)
Spiking of standards in DFU(11)
Roche Cobas c501®, cutoff 300 μg/LAmphetamines
II®
6.7 μg/mL
(m-CPP)
Native matrix(12)
Table I

False-Positive Results for Amphetamine Immunoassays

Cross-reacting drugImmunoassay platform (positive cutoff)Immunoassay nameLevel which false (+) occurredLevel of evidenceReference
BupropionSiemens Dimension®SYVA EMIT II Plus®OccurredRetrospective chart review(6)
ChlorpromazineRoche Hitachi 911® (1.0 μg/mL cutoff)SYVA EMIT-MAM®OccurredRetrospective chart review(1)
DMAARoche Modular P® (300 ng/mL cutoff)Roche kinetic interaction of microparticles in a solution
(KIMS®)
7.5 μg/mLSpiking of standards in DFU(7)
SYVA EMIT II Plus®3.1 μg/mLSpiking of standards in DFU(7)
Roche kinetic interaction of microparticles in a solution
(KIMS)® and SYVA EMIT II®
6.9 μg/mLTherapeutic dosing(7)
LabetalolNot mentionedNot mentionedOccurredCase report(8)
MetforminBiosite Triage®OccurredCase report(9)
OfloxacinTdxFlx® (300 ng/mL cutoff)AM/MA II®OccurredTherapeutic dosing(10)
PromethazineRoche Hitachi 911® (1.0 μg/mL cutoff)SYVA EMIT-MAM®OccurredRetrospective chart review(1)
Trazodone (m-CPP)Olympus UA 400®SYVA Ecstasy
EMIT II®
3 μg/mL
(m-CPP)
Spiking of standards in DFU(11)
Roche Cobas c501®, cutoff 300 μg/LAmphetamines
II®
6.7 μg/mL
(m-CPP)
Native matrix(12)
Cross-reacting drugImmunoassay platform (positive cutoff)Immunoassay nameLevel which false (+) occurredLevel of evidenceReference
BupropionSiemens Dimension®SYVA EMIT II Plus®OccurredRetrospective chart review(6)
ChlorpromazineRoche Hitachi 911® (1.0 μg/mL cutoff)SYVA EMIT-MAM®OccurredRetrospective chart review(1)
DMAARoche Modular P® (300 ng/mL cutoff)Roche kinetic interaction of microparticles in a solution
(KIMS®)
7.5 μg/mLSpiking of standards in DFU(7)
SYVA EMIT II Plus®3.1 μg/mLSpiking of standards in DFU(7)
Roche kinetic interaction of microparticles in a solution
(KIMS)® and SYVA EMIT II®
6.9 μg/mLTherapeutic dosing(7)
LabetalolNot mentionedNot mentionedOccurredCase report(8)
MetforminBiosite Triage®OccurredCase report(9)
OfloxacinTdxFlx® (300 ng/mL cutoff)AM/MA II®OccurredTherapeutic dosing(10)
PromethazineRoche Hitachi 911® (1.0 μg/mL cutoff)SYVA EMIT-MAM®OccurredRetrospective chart review(1)
Trazodone (m-CPP)Olympus UA 400®SYVA Ecstasy
EMIT II®
3 μg/mL
(m-CPP)
Spiking of standards in DFU(11)
Roche Cobas c501®, cutoff 300 μg/LAmphetamines
II®
6.7 μg/mL
(m-CPP)
Native matrix(12)
Figure 1.

Structures of methamphetamine, imipramine and diazepam (top) along with cross-reacting compounds (bottom).

Dimethylamylamine (DMAA), a widely used energy supplement, causes false-positive amphetamine screens in both the Roche KIMS® and SYVA EMIT II® kits on a Roche/Hitachi Modular P platform (7). With spiking experiments, Vorce et al. demonstrated that concentrations of 3.1 μg/mL of DMAA (4) analyzed on the EMIT II® assay caused a false-positive result for amphetamines (Figure 1). The KIMS® assay was more specific, requiring concentrations of 7.5 μg/mL and higher for a false-positive result. A separate examination was completed using urine samples from patients taking DMAA. When samples from patients taking DMAA were tested, concentrations at or >6.9 μg/mL returned false-positive results for both kits (Table I) (7).

Yee and Wu (8) described three reports of women prescribed labetalol testing positive for amphetamines. GC–MS analysis of these specimens failed to detect amphetamine or methamphetamine. The authors cite reports of a metabolite of labetalol, 3-amino-1-phenylbutane being known to cross-react with multiple amphetamine immunoassays (8).

Bupropion has also been shown to cross-react with amphetamine immunoassays. Casey et al. (6) completed a retrospective chart review of positive UDS amphetamine results which failed to confirm by GC–MS. Forty-one percent of these samples (53 patients) were from patients taking bupropion at the time of the UDS. After statistically ruling out poly-substance abuse as a cause, the investigators concluded that a large portion of these false-positive outcomes stemmed directly from the patients' bupropion intake (6). An important limitation of this study was the lack of spiking experiments demonstrating bupropion (or associated metabolites) truly causing the false-positive result (Table I).

An investigation by Melanson et al. (1) reveals that the structurally similar medications promethazine and chlorpromazine, both used to treat psychiatric conditions, can account for false-positive amphetamine UDSs . All patients with promethazine or chlorpromazine detected in their serum by liquid chromatography–photo-diode array (LC–PDA) detection, who also had an SYVA EMIT-MAM® (monoclonal amphetamine/methamphetamine) urine toxicology screen analyzed on the Roche Hitachi 911® platform, regardless of the results, were included in the study (Table I). A presumptive positive amphetamine result was considered if the sample was found to be above the cutoff concentration of the lowest calibrator for the assay, which is 1.0 μg/mL (1). Three urine specimens from patients taking promethazine were positive on the EMIT-MAM® assay, and all three were negative for amphetamines by LC–PDA. The authors also describe six false-positive amphetamine results in patients taking chlorpromazine that were negative for amphetamines via LC–PDA (Table I). The authors then analyzed all nine of these promethazine/chlorpromazine urine specimens on three other amphetamine UDS kits including SYVA EMIT-A®, Biosite® Triage and Agilent™ TesTcard 9. The results of the three promethazine and six chlorpromazine patient samples were all negative via the SYVA EMIT-A® and Biosite® Triage assays. However, the TesTcard 9 was subject to two false-positive results, one occurring from a patient taking promethazine and the other from a patient taking chlorpromazine (1). The authors conclude that both promethazine and chlorpromazine can cause false-positive amphetamines UDS and recommend confirmation of any presumptive positive results with a secondary method (1).

The commonly prescribed antidepressant, trazodone, has been reported to cause false positives in 3,4-methylenedioxymethamphetamine (MDMA) UDSs. A study conducted by Logan et al. (11) describes trazodone specifically cross-reacting with the EMIT II Plus® Ecstasy polyclonal assay, but not with the EMIT II Plus® Amphetamine monoclonal kit. The study consisted of spiking trazodone standards into DFU, then running the samples with both assay kits on the Olympus U400® platform (Table I). The authors concluded that trazodone can cause false-positive MDMA results at 3.0 μg/mL. The authors hypothesize that this false-positive result is most likely due to a metabolite of trazodone, meta-chlorophenylpiperazine (m-CPP), owing to its structural similarity with MDMA (11).

Baron et al. performed dilution experiments of m-CPP and trazodone standards in water and showed that false positives also occurred on the Amphetamine II assay on the Roche Cobas c501®. They demonstrated that a positive result for amphetamines will occur at concentrations of m-CPP at or >6.7 μg/mL when using a 1.0 μg/mL d-methamphetamine as the cutoff calibrator. These authors also showed that the concentration of trazodone would have to be >200 μg/mL to yield the positive result (12). The authors then inspected six patients with confirmed positive serum trazodone results by LC–MS and compared the results with the findings of the UDSs run on the aforementioned assay. Three patients with a trazodone concentration of >1.0 μg/mL and a serum m-CPP concentration of >0.85 μg/mL returned a positive result for amphetamines with the Amphetamines II® method (12). These presumptive positives were then confirmed negative for amphetamines via GC–MS (Table I).

Nomier and AL-Huseini (10) demonstrated that the Amphetamine/Methamphetamine II® kit run on the TdxFlx® platform can generate false-positive amphetamine results in the presence of ofloxacin . Urine samples were collected before and after administration of two 200 mg ofloxacin doses in six healthy male volunteers at an interval of 12 h. The urine samples were then analyzed using the Amphetamine/Methamphetamine II® assay on the TdxFlx. No samples prior to ofloxacin administration yielded a positive result for amphetamines. After administration however, all six samples returned positive results for amphetamines, the lowest of which was almost four times the concentration of the calibrator cutoff of 300 ng/mL (Table I) (10).

Tricyclic antidepressants

TCAs are a class of drugs mainly used to treat anxiety, eating disorders and depression (13). As the name suggests this class of compounds is based on a three-ringed organic framework, which is further modified to obtain desired pharmacological effects. The TCA assays have historically had a high rate of false positives. Drugs and/or metabolites described at or prior to the year 2000 as causing false-positive TCA results include carbamazepine (14) and cyclobenzaprine (15). Our literature search of recent interferences revealed that quetiapine (5), usually prescribed as an atypical antipsychotic (16), was the only recent drug found to yield false-positive TCA results (Figure 1).

In one report, the investigators created dilutions of quetiapine tablets dissolved in water and demonstrated that false positives for TCAs were occurring at 7.0 μg/mL using the Microgenics® Tricyclics Serum Tox EIA immunoassay on the Roche™ Hitachi 911® platform (16). A similar study demonstrated that quetiapine caused false-positive TCA results with the Microgenics® immunoassay on the Beckman™ LX20® platform and with the Syva® RapidTest d.a.u.® point-of-care (POC) kit (Table II). False positives occurred at 10.0 and 100.0 μg/mL, respectively (17). These investigators also analyzed urine from a patient prescribed a therapeutic dose of quetiapine 4 h prior to sample collection. Analysis of quetiapine in the patient's urine sample by gas chromatography–nitrogen–phosphorous detector (GC–NPD) determined that it contained 0.2 μg/mL of the parent compound (Table II) (17). Both the Microgenics® and Syva® immunoassays returned positive TCA results for the quetiapine dosed patient. This disagreement in the concentration cutoff for false positives occurring between native matrices and samples diluted in water is likely explained by the quetiapine metabolites present in patient specimens. Based on these data, it appears that quetiapine metabolites interfere with immunoassays for TCAs to a greater extent than the parent drug (17). Upon visual inspection of the structure of quetiapine and structures of common TCAs, one can recognize the structural similarities between them (Figure 1). This close structural relationship between quetiapine and the TCA drug class has been maintained as the probable cause for the observed interferences (17, 18).

Table II

False-Positive Results for TCA Immunoassays

Cross-reacting drugImmunoassay platform
(positive cutoff)
Immunoassay nameLevel which false (+) occurredLevel of evidenceReference
QuetiapineRoche (Hitachi 911) (300 ng/mL cutoff of nortriptyline)Microgenics®7.0 μg/mLSpiking of dissolved tablets in H2O(16)
Beckman LX20Microgenics
Tricyclics
Serum Tox EIA®
10.0 μg/mLSpiking of dissolved tablets in DFU(17)
Beckman LX20Microgenics
Tricyclics
Serum Tox EIA®
0.2 μg/mLTherapeutic dosing(17)
SYVA RapidTest d.a.u.®100.0 μg/mLSpiking of dissolved tablets in DFU(17)
SYVA RapidTest d.a.u.®0.2 μg/mLTherapeutic dosing(17)
Biosite Triage®>1,000.0 μg/mLSpiking of dissolved tablets in DFU(17)
Biosite Triage®Did not occurTherapeutic dosing(17)
Bio-Rad
Tox/See®
OccurredCase report(18)
Cross-reacting drugImmunoassay platform
(positive cutoff)
Immunoassay nameLevel which false (+) occurredLevel of evidenceReference
QuetiapineRoche (Hitachi 911) (300 ng/mL cutoff of nortriptyline)Microgenics®7.0 μg/mLSpiking of dissolved tablets in H2O(16)
Beckman LX20Microgenics
Tricyclics
Serum Tox EIA®
10.0 μg/mLSpiking of dissolved tablets in DFU(17)
Beckman LX20Microgenics
Tricyclics
Serum Tox EIA®
0.2 μg/mLTherapeutic dosing(17)
SYVA RapidTest d.a.u.®100.0 μg/mLSpiking of dissolved tablets in DFU(17)
SYVA RapidTest d.a.u.®0.2 μg/mLTherapeutic dosing(17)
Biosite Triage®>1,000.0 μg/mLSpiking of dissolved tablets in DFU(17)
Biosite Triage®Did not occurTherapeutic dosing(17)
Bio-Rad
Tox/See®
OccurredCase report(18)
Table II

False-Positive Results for TCA Immunoassays

Cross-reacting drugImmunoassay platform
(positive cutoff)
Immunoassay nameLevel which false (+) occurredLevel of evidenceReference
QuetiapineRoche (Hitachi 911) (300 ng/mL cutoff of nortriptyline)Microgenics®7.0 μg/mLSpiking of dissolved tablets in H2O(16)
Beckman LX20Microgenics
Tricyclics
Serum Tox EIA®
10.0 μg/mLSpiking of dissolved tablets in DFU(17)
Beckman LX20Microgenics
Tricyclics
Serum Tox EIA®
0.2 μg/mLTherapeutic dosing(17)
SYVA RapidTest d.a.u.®100.0 μg/mLSpiking of dissolved tablets in DFU(17)
SYVA RapidTest d.a.u.®0.2 μg/mLTherapeutic dosing(17)
Biosite Triage®>1,000.0 μg/mLSpiking of dissolved tablets in DFU(17)
Biosite Triage®Did not occurTherapeutic dosing(17)
Bio-Rad
Tox/See®
OccurredCase report(18)
Cross-reacting drugImmunoassay platform
(positive cutoff)
Immunoassay nameLevel which false (+) occurredLevel of evidenceReference
QuetiapineRoche (Hitachi 911) (300 ng/mL cutoff of nortriptyline)Microgenics®7.0 μg/mLSpiking of dissolved tablets in H2O(16)
Beckman LX20Microgenics
Tricyclics
Serum Tox EIA®
10.0 μg/mLSpiking of dissolved tablets in DFU(17)
Beckman LX20Microgenics
Tricyclics
Serum Tox EIA®
0.2 μg/mLTherapeutic dosing(17)
SYVA RapidTest d.a.u.®100.0 μg/mLSpiking of dissolved tablets in DFU(17)
SYVA RapidTest d.a.u.®0.2 μg/mLTherapeutic dosing(17)
Biosite Triage®>1,000.0 μg/mLSpiking of dissolved tablets in DFU(17)
Biosite Triage®Did not occurTherapeutic dosing(17)
Bio-Rad
Tox/See®
OccurredCase report(18)

Benzodiazepines

Benzodiazepines are widely prescribed as anxiolytics and hypnotics and are commonly detected in UDS. Immunoassay-based drug monitoring programs for benzodiazepines are well established; however, like most immunoassays, they are subject to false-positive results (Table III). Conversely, immunoassays for benzodiazepines also suffer from clinically significant false-negative results. The false negatives are not discussed here, but it is important that providers are aware of this limitation (21). Efavirenz (EFV) (6), used for the treatment of HIV, has been shown to cross-react with Biosite Triage 8® kit for benzodiazepines (Figure 1) (19). The authors created a blinded study in which half of the patients (n = 50) were therapeutically dosed with EFV and the other half (n = 50) were not. Running the Biosite Triage® kit for benzodiazepines, 46 of the 50 patients prescribed EFV returned a positive result, whereas 47 of the 50 patients not prescribed EFV were negative for benzodiazepines. Two other POC immunoassays (Drug Control 008® and Drug Screen Multi 5®) along with LC–MS-MS confirmation were then performed on the presumptive positive samples, all of which returned negative results (19). The authors also conducted spiking experiments with standards of EFV and its oxidized metabolite, 8-OH EFV, in DFU and revealed that both compounds are responsible for the observed false-positive results (Table III) (19). EFV, 8-OH EFV and many benzodiazepines are similar in structure, which may explain some of the observed cross-reactivity (Figure 1).

Table III

False-Positive Results for Benzodiazepine Immunoassays

Cross-reacting drugImmunoassay platform (positive cutoff)Immunoassay nameLevel which false (+) occurredLevel of evidenceReference
EFVBiosite Triage 8®OccurredRetrospective chart review(19)
SertralineAbbot Architect and Abbot Aeroset®Not mentionedOccurredRetrospective chart review(20)
Cross-reacting drugImmunoassay platform (positive cutoff)Immunoassay nameLevel which false (+) occurredLevel of evidenceReference
EFVBiosite Triage 8®OccurredRetrospective chart review(19)
SertralineAbbot Architect and Abbot Aeroset®Not mentionedOccurredRetrospective chart review(20)
Table III

False-Positive Results for Benzodiazepine Immunoassays

Cross-reacting drugImmunoassay platform (positive cutoff)Immunoassay nameLevel which false (+) occurredLevel of evidenceReference
EFVBiosite Triage 8®OccurredRetrospective chart review(19)
SertralineAbbot Architect and Abbot Aeroset®Not mentionedOccurredRetrospective chart review(20)
Cross-reacting drugImmunoassay platform (positive cutoff)Immunoassay nameLevel which false (+) occurredLevel of evidenceReference
EFVBiosite Triage 8®OccurredRetrospective chart review(19)
SertralineAbbot Architect and Abbot Aeroset®Not mentionedOccurredRetrospective chart review(20)

False-positive benzodiazepine results may also originate from the commonly prescribed antidepressant sertraline (20). A retrospective chart review was performed for all specimens in a 2-year span (January 2007–December 2008) that screened positive on the Abbott Architect® and Aeroset® platforms, but were negative by confirmatory testing (20). These false positives were then cross-referenced with the patients' prescription records. The authors concluded that 26.5% (26 samples) had false-positive benzodiazepine results related to sertraline use (Table III) although no spiking experiments were performed to support these observations (20).

Cannabinoids

Immunoassays are widely used to screen urine samples for recent marijuana abuse by detecting 11-nor-Δ9-tetrahydrocannabinol-9-carboxylic acid (11-nor-Δ9-THC-9-COOH), the major urinary metabolite of Δ9-tetrahydrocannabinol (THC) (7) (Figure 2).
Figure 2.

Structures of THC, morphine and PCP (top) along with cross-reacting compounds (bottom).

Patients prescribed antiretroviral therapy with EFV (6) produce urine samples that screened positive for THC exposure, despite the absence of 11-nor-Δ9-THC-9-COOH (Table IV). This interference has been attributed to EFVs major metabolite, EFV 8-glucuronide (EFV-8-G). The other major urinary metabolite, 8-OH EFV, and EFV were found to not interfere (22). Three immunoassays showed false-positive results for cannabinoid screening including Microgenics Corporation (Cedia® Dau Multi-Level THC), Biosite Incorporated (Triage® TOX Drug Screen) and Immunalysis Corporation [Cannabinoids (THCA/CTHC) Direct ELISA Kit] (22). Oosthuizen and Laurens (23) evaluated two POC devices (THC One Step Marijuana and Rapid Response Drugs-of-Abuse Test Strips) and two automated immunoassays (Roche Diagnostics Cannabinoids II and Beckman Coulter SYNCHRON Systems THC2) for THC false-positive results caused by EFV. They reported that the Rapid Response test strips yielded positive results in 28 of 30 patients taking 600 mg EFV therapy for at least 2 weeks, but the results from all of the other platforms were negative (Table IV) (23).

Table IV

False-Positive Results for Cannabinoid Immunoassays

Cross-reacting drugImmunoassay platform (positive cutoff)Immunoassay manufacturerLevel which false (+) occurredLevel of evidenceReference
EfavirenzCedia® Dau Multi-Level THC, Triage® TOX Drug Screen and Cannabinoids (THCA/CTHC) Direct ELISA Kit, cutoff 50 µg/LMicrogenics, BioSite and ImmunalysisOccurredCase report(22)
Rapid response drugs-of-abuse test strips, cutoff 50 ng/mLBTNXOccurredCase report(23)
IbuprofenEMIT® d.a.u. cutoff 20 µg/LSyvaOccurredCase report(24)
NaproxenEMIT® d.a.u. cutoff 20 µg/LSyvaOccurredCase report(24)
Niflumic acidKinetic interaction of microparticles in a solution (KIMS®), cutoff 50 ng/mLRoche2.5 μg/mLSpiking of standards in DFU(25)
Cross-reacting drugImmunoassay platform (positive cutoff)Immunoassay manufacturerLevel which false (+) occurredLevel of evidenceReference
EfavirenzCedia® Dau Multi-Level THC, Triage® TOX Drug Screen and Cannabinoids (THCA/CTHC) Direct ELISA Kit, cutoff 50 µg/LMicrogenics, BioSite and ImmunalysisOccurredCase report(22)
Rapid response drugs-of-abuse test strips, cutoff 50 ng/mLBTNXOccurredCase report(23)
IbuprofenEMIT® d.a.u. cutoff 20 µg/LSyvaOccurredCase report(24)
NaproxenEMIT® d.a.u. cutoff 20 µg/LSyvaOccurredCase report(24)
Niflumic acidKinetic interaction of microparticles in a solution (KIMS®), cutoff 50 ng/mLRoche2.5 μg/mLSpiking of standards in DFU(25)
Table IV

False-Positive Results for Cannabinoid Immunoassays

Cross-reacting drugImmunoassay platform (positive cutoff)Immunoassay manufacturerLevel which false (+) occurredLevel of evidenceReference
EfavirenzCedia® Dau Multi-Level THC, Triage® TOX Drug Screen and Cannabinoids (THCA/CTHC) Direct ELISA Kit, cutoff 50 µg/LMicrogenics, BioSite and ImmunalysisOccurredCase report(22)
Rapid response drugs-of-abuse test strips, cutoff 50 ng/mLBTNXOccurredCase report(23)
IbuprofenEMIT® d.a.u. cutoff 20 µg/LSyvaOccurredCase report(24)
NaproxenEMIT® d.a.u. cutoff 20 µg/LSyvaOccurredCase report(24)
Niflumic acidKinetic interaction of microparticles in a solution (KIMS®), cutoff 50 ng/mLRoche2.5 μg/mLSpiking of standards in DFU(25)
Cross-reacting drugImmunoassay platform (positive cutoff)Immunoassay manufacturerLevel which false (+) occurredLevel of evidenceReference
EfavirenzCedia® Dau Multi-Level THC, Triage® TOX Drug Screen and Cannabinoids (THCA/CTHC) Direct ELISA Kit, cutoff 50 µg/LMicrogenics, BioSite and ImmunalysisOccurredCase report(22)
Rapid response drugs-of-abuse test strips, cutoff 50 ng/mLBTNXOccurredCase report(23)
IbuprofenEMIT® d.a.u. cutoff 20 µg/LSyvaOccurredCase report(24)
NaproxenEMIT® d.a.u. cutoff 20 µg/LSyvaOccurredCase report(24)
Niflumic acidKinetic interaction of microparticles in a solution (KIMS®), cutoff 50 ng/mLRoche2.5 μg/mLSpiking of standards in DFU(25)

Niflumic acid (10) is a nonsteroidal, anti-inflammatory drug that inhibits cyclooxygenase-2 (Figure 2). A patient dosing study revealed that all 55 urine samples from five volunteers taking niflumic acid returned a positive THC result when analyzed by the KIMS® assay but were negative when analyzed by the EMIT THC® assay (25). Niflumic acid standards were added to DFU at 13 concentrations ranging from 1.25 to 1,000 μg/mL. A concentration of niflumic acid at ∼2.5 μg/mL had a KIMS® response equal to the 50 ng/mL cutoff (Table IV) (25).

Cotten et al. (26) identified several soap-based products that could potentially cause a false positive with the Vitros THC® (Cannabinoid) immunoassay on the Vitros 5600 (Ortho Clinical Diagnostics, Inc., Rochester, NY, USA). Four commercial baby soaps caused assay interference sufficient to yield a positive screen result (cutoff 20 ng/mL) (26). Despite the identification of several surfactant components contributing to the interference, the authors maintain that the exact mechanism for the false-positive results remains unclear (26).

Opiates

Opiates belong to a large class of compounds characterized by their ability to interact with endogenous opiate receptors (27). Opiate immunoassays typically target morphine (8) (Figure 2) and codeine, both naturally isolated from the opium poppy (Papaver somniferum). Semisynthetic opiates are similar in structure to morphine, whereas synthetic opioids that bind to opiate receptors generally require separate immunoassays for screening purposes.

Naloxone cross-reactivity with opiate immunoassays has been reported. Urine samples supplemented with 6.1 µg/mL of naloxone or greater were ‘opiate positive’ with the opiate CEDIA® assay (28). This finding is of particular importance as most patients with a suspected opiate overdose are treated with naloxone (Table V) (28). Buprenorphine (11) is a semisynthetic opioid (Figure 2) commonly administered in opiate agonist therapy to manage the patients dependent on opioids (48). When the Microgenics buprenorphine CEDIA® assay was evaluated for cross-reactivity, Pavlic et al. (30) demonstrated that 0.12 mg/mL morphine, 0.32 mg/mL methadone, 0.03 mg/mL codeine, 0.06 mg/mL dihydrocodeine and 0.52 mg/mL morphine-3-glucuronide produced positive results. This observed cross-reactivity is not unexpected as the structural characteristics of buprenorphine are similar to natural opiate derivatives (Table V). Patients are frequently tested simultaneously for buprenorphine compliance and avoidance of illicit opioids as a condition of continuance with opioid substitution therapy. It is particularly troublesome that a false-positive buprenorphine screening test could be caused by heroin use, which potentially yields large quantities of morphine in urine.

Table V

False-Positive Results for Opiate Immunoassays

Cross-reacting drugImmunoassay platform (positive cutoff)Immunoassay manufacturerLevel which false (+) occurredLevel of evidenceReference
Amisulpride and sulpirideCEDIA buprenorphine assay®, cutoff 5 ng/mLThermo Fisher Scientific130 mg/L for amisulpride and 250 mg/L for sulpirideSpiking of standards in DFU(29)
Codeine, dihydrocodeine, morphine, methadone and morphine-3-glucuronideCEDIA buprenorphine assay®, cutoff 5 ng/mLMicrogenics30 mg/L (codeine), 60 mg/L (dihydrocodeine), 120 mg/L (morphine), 320 mg/L (methadone) and 520 mg/L (morphine-3-glucuronide)Spiking of standards in DFU(30)
CreatinineEMIT® d.a.u. methadone, cutoff 300 ng/mLSyvaOccurredCase report(31)
DihydrocodeineCEDIA buprenorphine assay®, cutoff 5 ng/mLMicrogenicsOccurredCase report(32)
DiphenhydramineLateral-flow immunoassay for methadone, cutoff 300 ng/mLOne Step Multi-Drug Screen (ACON Laboratories)100 μg/mLSpiking of standards in DFU(33)
EMIT II® propoxyphene, cutoff 300 ng/mLSyva200 μg/LSpiking of standards in DFU(34)
Levofloxacine, ofloxacineAxSYM fluorescence polarization immunoassay morphine, cutoff 300 ng/mLAbbott1,700 μg/mLSpiking of standards in DFU(35)
Levofloxacine, ofloxacine, pefloxacine, enoxacineCEDIA morphine®, cutoff 300 ng/mLMicrogenics60–1,700 μg/mLSpiking of standards in DFU(35)
Levofloxacine, ofloxacine, pefloxacine, enoxacine, gatifloxacineEMIT II® morphine, cutoff 300 ng/mLSyva140–600 μg/mLSpiking of standards in DFU(35)
Levofloxacine, ofloxacine, pefloxacine, lomefloxacine, moxifloxacine, ciprofloxacine, norfloxacineAbuscreen OnLine morphine, cutoff 300 ng/mLRoche200–1,700 μg/mLSpiking of standards in DFU(35)
MorphineCEDIA buprenorphine assay®, cutoff 10 ng/mLMicrogenicsOccurredCase report(36)
NaloxoneCEDIA opiate®, cutoff 300 ng/mLMicrogenics6,100 μg/LSpiking of standards in DFU(28)
OfloxacinFluorescence immunoassay for opiatesTriage TM and Multi-5MOccurredCase report(37)
EMIT II® opiates, cutoff 300 µg/LSyva200 mg/LSpiking of standards in DFU(38)
PentazocineCedia heroin metabolite (6-AM) assay cutoff 10 ng/mLMicrogenicsOccurredCase report(39)
Psychotropic drugsKinetic interaction of microparticles in a solution
(KIMS®) methadone, cutoff 300 ng/mL
Roche Integra 8008 mg/L cyamemazine, 57 mg/L
alimemazine, 5 mg/L levomepromazine, 20 mg/L chlorpromazine,
100 mg/L clomipramine and 100 mg/L thioridazine
Spiking of standards in DFU(40)
QuetiapineKinetic interaction of microparticles
in solution (KIMS®) methadone cutoff 300 ng/mL
Roche COBAS Integra Methadone II test kitOccurredCase report(4143)
RifampicinKIMS® opiates (kinetic interaction of microparticles
in solution)
Cobas Integra156–5,000 µg/LSpiking of standards in DFU(44)
TapentadolDRI methadone enzyme immunoassay cutoff 130 ng/mLMicrogenics6,500 ng/mLSpiking of standards in DFU(45)
TramadolCEDIA buprenorphine assay® cutoff of 5 ng/mLMicrogenics100 mg/LStandard solutions of tramadol(46)
Lateral-flow
immunochrom buprenorphine assay, cutoff 5 ng/mL
Rapid One Buprenorphine
Test Cards (Am. Bio Medica)
50 mg/LStandard solutions of
tramadol
(46)
Lateral-flow
immunochromatographic buprenorphine assay, cutoff 5 ng/mL
QuikStrip OneStep Buprenorphine Test strips (CLIAwaived)OccurredCase report(46)
VerapamilMethadone, cutoff 300 ng/mLDiagnostic Reagents, Inc. (DRI)20 mg/LSpiking of standards in DFU(47)
Cross-reacting drugImmunoassay platform (positive cutoff)Immunoassay manufacturerLevel which false (+) occurredLevel of evidenceReference
Amisulpride and sulpirideCEDIA buprenorphine assay®, cutoff 5 ng/mLThermo Fisher Scientific130 mg/L for amisulpride and 250 mg/L for sulpirideSpiking of standards in DFU(29)
Codeine, dihydrocodeine, morphine, methadone and morphine-3-glucuronideCEDIA buprenorphine assay®, cutoff 5 ng/mLMicrogenics30 mg/L (codeine), 60 mg/L (dihydrocodeine), 120 mg/L (morphine), 320 mg/L (methadone) and 520 mg/L (morphine-3-glucuronide)Spiking of standards in DFU(30)
CreatinineEMIT® d.a.u. methadone, cutoff 300 ng/mLSyvaOccurredCase report(31)
DihydrocodeineCEDIA buprenorphine assay®, cutoff 5 ng/mLMicrogenicsOccurredCase report(32)
DiphenhydramineLateral-flow immunoassay for methadone, cutoff 300 ng/mLOne Step Multi-Drug Screen (ACON Laboratories)100 μg/mLSpiking of standards in DFU(33)
EMIT II® propoxyphene, cutoff 300 ng/mLSyva200 μg/LSpiking of standards in DFU(34)
Levofloxacine, ofloxacineAxSYM fluorescence polarization immunoassay morphine, cutoff 300 ng/mLAbbott1,700 μg/mLSpiking of standards in DFU(35)
Levofloxacine, ofloxacine, pefloxacine, enoxacineCEDIA morphine®, cutoff 300 ng/mLMicrogenics60–1,700 μg/mLSpiking of standards in DFU(35)
Levofloxacine, ofloxacine, pefloxacine, enoxacine, gatifloxacineEMIT II® morphine, cutoff 300 ng/mLSyva140–600 μg/mLSpiking of standards in DFU(35)
Levofloxacine, ofloxacine, pefloxacine, lomefloxacine, moxifloxacine, ciprofloxacine, norfloxacineAbuscreen OnLine morphine, cutoff 300 ng/mLRoche200–1,700 μg/mLSpiking of standards in DFU(35)
MorphineCEDIA buprenorphine assay®, cutoff 10 ng/mLMicrogenicsOccurredCase report(36)
NaloxoneCEDIA opiate®, cutoff 300 ng/mLMicrogenics6,100 μg/LSpiking of standards in DFU(28)
OfloxacinFluorescence immunoassay for opiatesTriage TM and Multi-5MOccurredCase report(37)
EMIT II® opiates, cutoff 300 µg/LSyva200 mg/LSpiking of standards in DFU(38)
PentazocineCedia heroin metabolite (6-AM) assay cutoff 10 ng/mLMicrogenicsOccurredCase report(39)
Psychotropic drugsKinetic interaction of microparticles in a solution
(KIMS®) methadone, cutoff 300 ng/mL
Roche Integra 8008 mg/L cyamemazine, 57 mg/L
alimemazine, 5 mg/L levomepromazine, 20 mg/L chlorpromazine,
100 mg/L clomipramine and 100 mg/L thioridazine
Spiking of standards in DFU(40)
QuetiapineKinetic interaction of microparticles
in solution (KIMS®) methadone cutoff 300 ng/mL
Roche COBAS Integra Methadone II test kitOccurredCase report(4143)
RifampicinKIMS® opiates (kinetic interaction of microparticles
in solution)
Cobas Integra156–5,000 µg/LSpiking of standards in DFU(44)
TapentadolDRI methadone enzyme immunoassay cutoff 130 ng/mLMicrogenics6,500 ng/mLSpiking of standards in DFU(45)
TramadolCEDIA buprenorphine assay® cutoff of 5 ng/mLMicrogenics100 mg/LStandard solutions of tramadol(46)
Lateral-flow
immunochrom buprenorphine assay, cutoff 5 ng/mL
Rapid One Buprenorphine
Test Cards (Am. Bio Medica)
50 mg/LStandard solutions of
tramadol
(46)
Lateral-flow
immunochromatographic buprenorphine assay, cutoff 5 ng/mL
QuikStrip OneStep Buprenorphine Test strips (CLIAwaived)OccurredCase report(46)
VerapamilMethadone, cutoff 300 ng/mLDiagnostic Reagents, Inc. (DRI)20 mg/LSpiking of standards in DFU(47)
Table V

False-Positive Results for Opiate Immunoassays

Cross-reacting drugImmunoassay platform (positive cutoff)Immunoassay manufacturerLevel which false (+) occurredLevel of evidenceReference
Amisulpride and sulpirideCEDIA buprenorphine assay®, cutoff 5 ng/mLThermo Fisher Scientific130 mg/L for amisulpride and 250 mg/L for sulpirideSpiking of standards in DFU(29)
Codeine, dihydrocodeine, morphine, methadone and morphine-3-glucuronideCEDIA buprenorphine assay®, cutoff 5 ng/mLMicrogenics30 mg/L (codeine), 60 mg/L (dihydrocodeine), 120 mg/L (morphine), 320 mg/L (methadone) and 520 mg/L (morphine-3-glucuronide)Spiking of standards in DFU(30)
CreatinineEMIT® d.a.u. methadone, cutoff 300 ng/mLSyvaOccurredCase report(31)
DihydrocodeineCEDIA buprenorphine assay®, cutoff 5 ng/mLMicrogenicsOccurredCase report(32)
DiphenhydramineLateral-flow immunoassay for methadone, cutoff 300 ng/mLOne Step Multi-Drug Screen (ACON Laboratories)100 μg/mLSpiking of standards in DFU(33)
EMIT II® propoxyphene, cutoff 300 ng/mLSyva200 μg/LSpiking of standards in DFU(34)
Levofloxacine, ofloxacineAxSYM fluorescence polarization immunoassay morphine, cutoff 300 ng/mLAbbott1,700 μg/mLSpiking of standards in DFU(35)
Levofloxacine, ofloxacine, pefloxacine, enoxacineCEDIA morphine®, cutoff 300 ng/mLMicrogenics60–1,700 μg/mLSpiking of standards in DFU(35)
Levofloxacine, ofloxacine, pefloxacine, enoxacine, gatifloxacineEMIT II® morphine, cutoff 300 ng/mLSyva140–600 μg/mLSpiking of standards in DFU(35)
Levofloxacine, ofloxacine, pefloxacine, lomefloxacine, moxifloxacine, ciprofloxacine, norfloxacineAbuscreen OnLine morphine, cutoff 300 ng/mLRoche200–1,700 μg/mLSpiking of standards in DFU(35)
MorphineCEDIA buprenorphine assay®, cutoff 10 ng/mLMicrogenicsOccurredCase report(36)
NaloxoneCEDIA opiate®, cutoff 300 ng/mLMicrogenics6,100 μg/LSpiking of standards in DFU(28)
OfloxacinFluorescence immunoassay for opiatesTriage TM and Multi-5MOccurredCase report(37)
EMIT II® opiates, cutoff 300 µg/LSyva200 mg/LSpiking of standards in DFU(38)
PentazocineCedia heroin metabolite (6-AM) assay cutoff 10 ng/mLMicrogenicsOccurredCase report(39)
Psychotropic drugsKinetic interaction of microparticles in a solution
(KIMS®) methadone, cutoff 300 ng/mL
Roche Integra 8008 mg/L cyamemazine, 57 mg/L
alimemazine, 5 mg/L levomepromazine, 20 mg/L chlorpromazine,
100 mg/L clomipramine and 100 mg/L thioridazine
Spiking of standards in DFU(40)
QuetiapineKinetic interaction of microparticles
in solution (KIMS®) methadone cutoff 300 ng/mL
Roche COBAS Integra Methadone II test kitOccurredCase report(4143)
RifampicinKIMS® opiates (kinetic interaction of microparticles
in solution)
Cobas Integra156–5,000 µg/LSpiking of standards in DFU(44)
TapentadolDRI methadone enzyme immunoassay cutoff 130 ng/mLMicrogenics6,500 ng/mLSpiking of standards in DFU(45)
TramadolCEDIA buprenorphine assay® cutoff of 5 ng/mLMicrogenics100 mg/LStandard solutions of tramadol(46)
Lateral-flow
immunochrom buprenorphine assay, cutoff 5 ng/mL
Rapid One Buprenorphine
Test Cards (Am. Bio Medica)
50 mg/LStandard solutions of
tramadol
(46)
Lateral-flow
immunochromatographic buprenorphine assay, cutoff 5 ng/mL
QuikStrip OneStep Buprenorphine Test strips (CLIAwaived)OccurredCase report(46)
VerapamilMethadone, cutoff 300 ng/mLDiagnostic Reagents, Inc. (DRI)20 mg/LSpiking of standards in DFU(47)
Cross-reacting drugImmunoassay platform (positive cutoff)Immunoassay manufacturerLevel which false (+) occurredLevel of evidenceReference
Amisulpride and sulpirideCEDIA buprenorphine assay®, cutoff 5 ng/mLThermo Fisher Scientific130 mg/L for amisulpride and 250 mg/L for sulpirideSpiking of standards in DFU(29)
Codeine, dihydrocodeine, morphine, methadone and morphine-3-glucuronideCEDIA buprenorphine assay®, cutoff 5 ng/mLMicrogenics30 mg/L (codeine), 60 mg/L (dihydrocodeine), 120 mg/L (morphine), 320 mg/L (methadone) and 520 mg/L (morphine-3-glucuronide)Spiking of standards in DFU(30)
CreatinineEMIT® d.a.u. methadone, cutoff 300 ng/mLSyvaOccurredCase report(31)
DihydrocodeineCEDIA buprenorphine assay®, cutoff 5 ng/mLMicrogenicsOccurredCase report(32)
DiphenhydramineLateral-flow immunoassay for methadone, cutoff 300 ng/mLOne Step Multi-Drug Screen (ACON Laboratories)100 μg/mLSpiking of standards in DFU(33)
EMIT II® propoxyphene, cutoff 300 ng/mLSyva200 μg/LSpiking of standards in DFU(34)
Levofloxacine, ofloxacineAxSYM fluorescence polarization immunoassay morphine, cutoff 300 ng/mLAbbott1,700 μg/mLSpiking of standards in DFU(35)
Levofloxacine, ofloxacine, pefloxacine, enoxacineCEDIA morphine®, cutoff 300 ng/mLMicrogenics60–1,700 μg/mLSpiking of standards in DFU(35)
Levofloxacine, ofloxacine, pefloxacine, enoxacine, gatifloxacineEMIT II® morphine, cutoff 300 ng/mLSyva140–600 μg/mLSpiking of standards in DFU(35)
Levofloxacine, ofloxacine, pefloxacine, lomefloxacine, moxifloxacine, ciprofloxacine, norfloxacineAbuscreen OnLine morphine, cutoff 300 ng/mLRoche200–1,700 μg/mLSpiking of standards in DFU(35)
MorphineCEDIA buprenorphine assay®, cutoff 10 ng/mLMicrogenicsOccurredCase report(36)
NaloxoneCEDIA opiate®, cutoff 300 ng/mLMicrogenics6,100 μg/LSpiking of standards in DFU(28)
OfloxacinFluorescence immunoassay for opiatesTriage TM and Multi-5MOccurredCase report(37)
EMIT II® opiates, cutoff 300 µg/LSyva200 mg/LSpiking of standards in DFU(38)
PentazocineCedia heroin metabolite (6-AM) assay cutoff 10 ng/mLMicrogenicsOccurredCase report(39)
Psychotropic drugsKinetic interaction of microparticles in a solution
(KIMS®) methadone, cutoff 300 ng/mL
Roche Integra 8008 mg/L cyamemazine, 57 mg/L
alimemazine, 5 mg/L levomepromazine, 20 mg/L chlorpromazine,
100 mg/L clomipramine and 100 mg/L thioridazine
Spiking of standards in DFU(40)
QuetiapineKinetic interaction of microparticles
in solution (KIMS®) methadone cutoff 300 ng/mL
Roche COBAS Integra Methadone II test kitOccurredCase report(4143)
RifampicinKIMS® opiates (kinetic interaction of microparticles
in solution)
Cobas Integra156–5,000 µg/LSpiking of standards in DFU(44)
TapentadolDRI methadone enzyme immunoassay cutoff 130 ng/mLMicrogenics6,500 ng/mLSpiking of standards in DFU(45)
TramadolCEDIA buprenorphine assay® cutoff of 5 ng/mLMicrogenics100 mg/LStandard solutions of tramadol(46)
Lateral-flow
immunochrom buprenorphine assay, cutoff 5 ng/mL
Rapid One Buprenorphine
Test Cards (Am. Bio Medica)
50 mg/LStandard solutions of
tramadol
(46)
Lateral-flow
immunochromatographic buprenorphine assay, cutoff 5 ng/mL
QuikStrip OneStep Buprenorphine Test strips (CLIAwaived)OccurredCase report(46)
VerapamilMethadone, cutoff 300 ng/mLDiagnostic Reagents, Inc. (DRI)20 mg/LSpiking of standards in DFU(47)

The buprenorphine assay is also subject to cross-reactivity from the synthetic opioid tramadol (Table V) (46). Tramadol interference with the CEDIA® buprenorphine assay is clinically relevant if a 5 ng/mL calibration cutoff is used. However, when the cutoff was increased from 5 to 20 ng/mL, all false positives were eliminated (Table V) (46). Amisulpride and sulpiride were also found to cross-react with the CEDIA® buprenorphine assay. Although the cross-reactivity is low, it remains significant given the high concentrations of these compounds that are commonly found in urine relative to the low cutoff (5 μg/L) of the buprenorphine immunoassay (29).

Methadone, a synthetic opioid, is used medically as an analgesic for the management of severe pain and in maintenance doses for patients with opioid dependency. Several reports exist regarding false-positive results for methadone screening for patients prescribed quetiapine (41, 42). Cherwinski et al. (49) reported that 125 mg/day doses of quetiapine were sufficient to produce a false positive of methadone screening results. Quetiapine is a widely used antipsychotic and is structurally similar to methadone and is maintained as the probable cause for the observed cross-reactivity (Table V) (43).

Miscellaneous drugs

This section is a compilation of drugs of abuse that do not fit into a particular class and are discussed individually. Phencyclidine, commonly known as PCP (9), is a synthetic, dissociative anesthetic, which has been sporadically abused for the last 50 years (Figure 2) (50). Many immunoassays that are available purport to specifically identify PCP in urine specimens. However, the low prevalence of PCP use combined with the low specificity of PCP immunoassays makes the positive predictive value of PCP screen very low (Table VI).

Table VI

False-Positive Results for Miscellaneous Drug (PCP, LSD and Barbiturates) Immunoassays

Immunoassay platform (positive cutoff)Immunoassay manufacturerLevel which false (+) occurredLevel of evidenceReference
Cross-reacting drug for PCP
 LamotrigineBio-Rad Tox/See®7.6 μg/mL
(serum)
Case report(50)
 MDPVBeckman Coulter DxC 800 ®, cutoff 25 ng/mLSynchron®3.1 μg/mLSpiking of standards in DFU(51)
 TramadolNot definedSyva EMIT II Plus®5.4 μg/mLCase report(52)
Siemens ADVIA 1800®DRI PCP
Microgenics®, cutoff 25 ng/mL
500 μg/mL at lowest cutoff 25 ng/mLSpiking of standards in DFU(53)
Not definedSyva EMIT II®OccurredCase report(54)
Siemens Dimension®Syva EMIT II®OccurredRetrospective chart review(3)
 VenlafaxineSyva RapidTest d.a.u.®, cutoff 25 ng/mL200 μg/mLSpiking of standards in DFU(55)
Abbot AxSYM®FPIA for PCPOccurred cutoff 25 ng/mLCase report(56)
Instant-View Multi-Drug Screen®OccurredCase report(57)
Cross-reacting drug for LSD
 AmbroxolCEDIA DAU LSD assay cutoff 0.5 ng/mLHitachi 917-analyzer3 mg/LSpiking of standards in DFU(58)
 Amitriptyline, benzphetamine, bupropion, buspirone, cephradine, chlorpromazine, desipramine, diltiazem, doxepin, fentanyl, fluoxetine, haloperidol, imipramine, labetalol, metoclopramide, prochlorperazine, risperidone, sertraline, thioridazine, trazodone and verapamilEmit II LSD® assay from Behring diagnostics cutoff 0.5 ng/mLSyva-30R automated assayVariable (0.1–<1,000 mg/L)Spiking of standards in DFU(59)
 Ergonovine, lysergol, brompheniramine maleate, imipramine HCl and methylphenidate HClEMIT II LSD® cutoff 500 pg/mLDade Behring4,000 ng/mL (ergonovine), 25,000 ng/mL (lysergol and brompheniramine maleate), 50,000 ng/mL (imipramine HCl) and 50,000 ng/mL (methylphenidate HCl)Spiking of standards in DFU(60)
 FentanylCEDIA LSD® assay and Emit II LSD® assayMicrogenics and SYVA40 μg/LSpiking of standards in DFU(61)
 SertralineCEDIA®Microgenics1.5 mg/LSpiking of standards in DFU(62)
Cross-reacting drug for barbiturates
 IbuprofenFPIA (cutoff 500 μg/L)Abbott TDxOccurredCase report(24)
 NaproxenFPIA (cutoff 500 μg/L)Abbott TDxOccurredCase report(24)
Immunoassay platform (positive cutoff)Immunoassay manufacturerLevel which false (+) occurredLevel of evidenceReference
Cross-reacting drug for PCP
 LamotrigineBio-Rad Tox/See®7.6 μg/mL
(serum)
Case report(50)
 MDPVBeckman Coulter DxC 800 ®, cutoff 25 ng/mLSynchron®3.1 μg/mLSpiking of standards in DFU(51)
 TramadolNot definedSyva EMIT II Plus®5.4 μg/mLCase report(52)
Siemens ADVIA 1800®DRI PCP
Microgenics®, cutoff 25 ng/mL
500 μg/mL at lowest cutoff 25 ng/mLSpiking of standards in DFU(53)
Not definedSyva EMIT II®OccurredCase report(54)
Siemens Dimension®Syva EMIT II®OccurredRetrospective chart review(3)
 VenlafaxineSyva RapidTest d.a.u.®, cutoff 25 ng/mL200 μg/mLSpiking of standards in DFU(55)
Abbot AxSYM®FPIA for PCPOccurred cutoff 25 ng/mLCase report(56)
Instant-View Multi-Drug Screen®OccurredCase report(57)
Cross-reacting drug for LSD
 AmbroxolCEDIA DAU LSD assay cutoff 0.5 ng/mLHitachi 917-analyzer3 mg/LSpiking of standards in DFU(58)
 Amitriptyline, benzphetamine, bupropion, buspirone, cephradine, chlorpromazine, desipramine, diltiazem, doxepin, fentanyl, fluoxetine, haloperidol, imipramine, labetalol, metoclopramide, prochlorperazine, risperidone, sertraline, thioridazine, trazodone and verapamilEmit II LSD® assay from Behring diagnostics cutoff 0.5 ng/mLSyva-30R automated assayVariable (0.1–<1,000 mg/L)Spiking of standards in DFU(59)
 Ergonovine, lysergol, brompheniramine maleate, imipramine HCl and methylphenidate HClEMIT II LSD® cutoff 500 pg/mLDade Behring4,000 ng/mL (ergonovine), 25,000 ng/mL (lysergol and brompheniramine maleate), 50,000 ng/mL (imipramine HCl) and 50,000 ng/mL (methylphenidate HCl)Spiking of standards in DFU(60)
 FentanylCEDIA LSD® assay and Emit II LSD® assayMicrogenics and SYVA40 μg/LSpiking of standards in DFU(61)
 SertralineCEDIA®Microgenics1.5 mg/LSpiking of standards in DFU(62)
Cross-reacting drug for barbiturates
 IbuprofenFPIA (cutoff 500 μg/L)Abbott TDxOccurredCase report(24)
 NaproxenFPIA (cutoff 500 μg/L)Abbott TDxOccurredCase report(24)
Table VI

False-Positive Results for Miscellaneous Drug (PCP, LSD and Barbiturates) Immunoassays

Immunoassay platform (positive cutoff)Immunoassay manufacturerLevel which false (+) occurredLevel of evidenceReference
Cross-reacting drug for PCP
 LamotrigineBio-Rad Tox/See®7.6 μg/mL
(serum)
Case report(50)
 MDPVBeckman Coulter DxC 800 ®, cutoff 25 ng/mLSynchron®3.1 μg/mLSpiking of standards in DFU(51)
 TramadolNot definedSyva EMIT II Plus®5.4 μg/mLCase report(52)
Siemens ADVIA 1800®DRI PCP
Microgenics®, cutoff 25 ng/mL
500 μg/mL at lowest cutoff 25 ng/mLSpiking of standards in DFU(53)
Not definedSyva EMIT II®OccurredCase report(54)
Siemens Dimension®Syva EMIT II®OccurredRetrospective chart review(3)
 VenlafaxineSyva RapidTest d.a.u.®, cutoff 25 ng/mL200 μg/mLSpiking of standards in DFU(55)
Abbot AxSYM®FPIA for PCPOccurred cutoff 25 ng/mLCase report(56)
Instant-View Multi-Drug Screen®OccurredCase report(57)
Cross-reacting drug for LSD
 AmbroxolCEDIA DAU LSD assay cutoff 0.5 ng/mLHitachi 917-analyzer3 mg/LSpiking of standards in DFU(58)
 Amitriptyline, benzphetamine, bupropion, buspirone, cephradine, chlorpromazine, desipramine, diltiazem, doxepin, fentanyl, fluoxetine, haloperidol, imipramine, labetalol, metoclopramide, prochlorperazine, risperidone, sertraline, thioridazine, trazodone and verapamilEmit II LSD® assay from Behring diagnostics cutoff 0.5 ng/mLSyva-30R automated assayVariable (0.1–<1,000 mg/L)Spiking of standards in DFU(59)
 Ergonovine, lysergol, brompheniramine maleate, imipramine HCl and methylphenidate HClEMIT II LSD® cutoff 500 pg/mLDade Behring4,000 ng/mL (ergonovine), 25,000 ng/mL (lysergol and brompheniramine maleate), 50,000 ng/mL (imipramine HCl) and 50,000 ng/mL (methylphenidate HCl)Spiking of standards in DFU(60)
 FentanylCEDIA LSD® assay and Emit II LSD® assayMicrogenics and SYVA40 μg/LSpiking of standards in DFU(61)
 SertralineCEDIA®Microgenics1.5 mg/LSpiking of standards in DFU(62)
Cross-reacting drug for barbiturates
 IbuprofenFPIA (cutoff 500 μg/L)Abbott TDxOccurredCase report(24)
 NaproxenFPIA (cutoff 500 μg/L)Abbott TDxOccurredCase report(24)
Immunoassay platform (positive cutoff)Immunoassay manufacturerLevel which false (+) occurredLevel of evidenceReference
Cross-reacting drug for PCP
 LamotrigineBio-Rad Tox/See®7.6 μg/mL
(serum)
Case report(50)
 MDPVBeckman Coulter DxC 800 ®, cutoff 25 ng/mLSynchron®3.1 μg/mLSpiking of standards in DFU(51)
 TramadolNot definedSyva EMIT II Plus®5.4 μg/mLCase report(52)
Siemens ADVIA 1800®DRI PCP
Microgenics®, cutoff 25 ng/mL
500 μg/mL at lowest cutoff 25 ng/mLSpiking of standards in DFU(53)
Not definedSyva EMIT II®OccurredCase report(54)
Siemens Dimension®Syva EMIT II®OccurredRetrospective chart review(3)
 VenlafaxineSyva RapidTest d.a.u.®, cutoff 25 ng/mL200 μg/mLSpiking of standards in DFU(55)
Abbot AxSYM®FPIA for PCPOccurred cutoff 25 ng/mLCase report(56)
Instant-View Multi-Drug Screen®OccurredCase report(57)
Cross-reacting drug for LSD
 AmbroxolCEDIA DAU LSD assay cutoff 0.5 ng/mLHitachi 917-analyzer3 mg/LSpiking of standards in DFU(58)
 Amitriptyline, benzphetamine, bupropion, buspirone, cephradine, chlorpromazine, desipramine, diltiazem, doxepin, fentanyl, fluoxetine, haloperidol, imipramine, labetalol, metoclopramide, prochlorperazine, risperidone, sertraline, thioridazine, trazodone and verapamilEmit II LSD® assay from Behring diagnostics cutoff 0.5 ng/mLSyva-30R automated assayVariable (0.1–<1,000 mg/L)Spiking of standards in DFU(59)
 Ergonovine, lysergol, brompheniramine maleate, imipramine HCl and methylphenidate HClEMIT II LSD® cutoff 500 pg/mLDade Behring4,000 ng/mL (ergonovine), 25,000 ng/mL (lysergol and brompheniramine maleate), 50,000 ng/mL (imipramine HCl) and 50,000 ng/mL (methylphenidate HCl)Spiking of standards in DFU(60)
 FentanylCEDIA LSD® assay and Emit II LSD® assayMicrogenics and SYVA40 μg/LSpiking of standards in DFU(61)
 SertralineCEDIA®Microgenics1.5 mg/LSpiking of standards in DFU(62)
Cross-reacting drug for barbiturates
 IbuprofenFPIA (cutoff 500 μg/L)Abbott TDxOccurredCase report(24)
 NaproxenFPIA (cutoff 500 μg/L)Abbott TDxOccurredCase report(24)

Venlafaxine (12), often prescribed as an antidepressant, has been shown to cause false-positive PCP results with multiple immunoassays (Figure 2). One case report presented by Bond et al. (56) describes a false-positive PCP immunoassay from a patient who overdosed on venlafaxine. A UDS performed using the Abbott AxSYM® platform was positive for PCP. Confirmative testing via GC–MS for PCP was negative. The only substance identified in the patient's serum via GC–MS was venlafaxine that was present at a concentration of >24 μg/mL, a concentration almost 1,000 times the therapeutic dose. The authors spiked DFU with an equivalent quantity of venlafaxine and its metabolite O-desmethylvenlafaxine and analyzed the samples using the Abbott AxSYM® platform. The spiked samples were positive for PCP even when diluted four times with DFU (56). Another instance of venlafaxine inference for PCP comes from a case report by Sena et al. Three separate patient urine specimens tested with the Syva RapidTest d.a.u.® method were positive for PCP using a cutoff of 25-ng/mL (Table VI) (55). Venlafaxine was found to be common in all three patients. The three samples were then retested on different immunoassay platforms for PCP. These included Syva Emit II Plus® PCP assay on the Roche Cobas c501® platform, the OnTrak TesTstik PCP® kit and the Biosite Triage 8®, all of which had a cutoff of 25 ng/mL and all of which returned negative results for PCP. GC–MS testing confirmed these negative PCP results for all three patients (Table VI) (55).

Tramadol also causes false-positive PCP immunoassay results. One study reports two cases of false-positive PCP urine screens from patients taking tramadol using the DRI PCP Microgenics® kit on the Siemens ADVIA 1800® platform (53). No PCP was detected in these specimens with GC–MS. The first patient had acknowledged to taking 10 times her prescribed tramadol dose each day for the 2 days prior to her urine samples being collected. A broad-spectrum GC–MS analysis was conducted on the patient's urine sample that showed a large and distorted peak for tramadol, indicating a high but undefined concentration of the drug. The second patient had also consumed a large quantity of tramadol (eighty-four 50 mg tramadol tablets) and showed similar GC–MS results. The authors performed dilution studies of tramadol, O-desmethyltramadol and N-desmethyltramadol standards in DFU and demonstrated that concentrations of 500 μg/mL for both tramadol and N-desmethyltramadol were found to trigger a positive PCP response at a cutoff of 25 ng/mL (53). False positives for PCP from tramadol overdose were also noted in two patients using the Syva EMIT II Plus® immunoassay (Table VI) (54). Both patients were admitted to the emergency department with PCP positive UDSs. Confirmatory analysis by GC–MS showed no detectable PCP. Tramadol, however, was found in high amounts although exact quantification was not established (54). The authors suggest that these PCP false positives are due to the two compounds' structural similarities (54).

3,4-Methylenedioxypyrovalerone (MDPV) is a synthetic cathinone found in illicit ‘bath salts’ and was revealed by Macher and Penders to cause false-positive PCP results. The case involved a patient who was admitted to the emergency department with an apparent overdose. A urine sample was collected and analyzed with a Beckman Coulter DxC 800® method run on a Synchron® platform (51). Confirmatory testing was not conducted on the patient's urine sample but was accomplished via GC–MS using a postmortem blood sample. The results indicated no PCP was present in the patient's blood at the time of death. MDPV, however, was detected in the blood at a concentration of 1.0 μg/mL, prompting further investigative studies (Table VI). Dilutions of MDPV standards in DFU were made and analyzed, which showed that an MDPV concentration of >3.1 μg/mL can cause false-positive results for PCP (51).

Immunoassays for the hallucinogen, lysergic acid diethylamide (LSD), have been shown to cross-react with a number of different drug types most notably when using the EMIT II® kit (59, 60). Twelve patients with unexpected positive results for LSD were caused by the mucolytic drug ambroxol (58). The compound also exhibits a significant cross-reactivity in the CEDIA DAU LSD® assay (58). More recently, the CEDIA LSD UDS® was shown to be subject to false positives from sertraline (62). The authors spiked DFU with sertraline (ranging from 0 to 2 μg/mL) (Table VI). The results obtained from the urine samples fortified with sertraline showed that the CEDIA LSD® screening was positive when concentrations reached 1.5 μg/mL (62).

Conclusion

This literature review summarizes many reports that demonstrate false-positive immunoassay results can be caused by a variety of compounds. Common structural motifs account for some of these occurrences, but false-positive results can also be caused by dissimilar structures. It is important to remember that antibodies are binding to a three-dimensional structure. While some similarities can be seen between two-dimensional chemical structures of cross-reacting compounds, the absence of a similar chemical appearance does not rule out immunoassay cross-reactivity. It is also important to recognize that drug metabolites can also cause false-positive results. Simply adding the parent drug to DFU and demonstrating the absence of a positive result does not rule out ingestion of a particular drug as the cause of a false-positive result. Immunoassays provide useful clinical information, but should be viewed as ‘presumptive positive’ results until confirmed by an independent chemical technique such as GC–MS or liquid chromatography–tandem mass spectrometry.

Acknowledgments

The authors acknowledge the generous funding provided by Waters Corporation for the University of California, San Diego (UCSD) clinical chemistry fellowship program. The authors graciously acknowledge Dr Judith A. Stone (UCSD) for her input in preparation of this manuscript.

References

1
Melanson
S.E.
Lee-Lewandrowski
E.
Griggs
D.A.
Long
W.H.
Flood
J.G.
Reduced interference by phenothiazines in amphetamine drug of abuse immunoassays
Archives of Pathology & Laboratory Medicine
2006
, vol. 
130
 (pg. 
1834
-
1838
)
2
Reisfield
G.M.
Goldberger
B.A.
Bertholf
R.L.
False-positive’ and ‘false-negative’ test results in clinical urine drug testing
Bioanalysis
2009
, vol. 
1
 (pg. 
937
-
952
)
3
Rengarajan
A.
Mullins
M.E.
How often do false-positive phencyclidine urine screens occur with use of common medications?
Clinical Toxicology (Philadelphia, PA.)
2013
, vol. 
51
 (pg. 
493
-
496
)
4
Brahm
N.C.
Yeager
L.L.
Fox
M.D.
Farmer
K.C.
Palmer
T.A.
Commonly prescribed medications and potential false-positive urine drug screens
American Journal of Health-System Pharmacy
2010
, vol. 
67
 (pg. 
1344
-
1350
)
5
Drevets
W.C.
Gautier
C.
Price
J.C.
Kupfer
D.J.
Kinahan
P.E.
Grace
A.A.
, et al. 
Amphetamine-induced dopamine release in human ventral striatum correlates with euphoria
Biological Psychiatry
2001
, vol. 
49
 (pg. 
81
-
96
)
6
Casey
E.R.
Scott
M.G.
Tang
S.
Mullins
M.E.
Frequency of false positive amphetamine screens due to bupropion using the Syva Emit II Immunoassay
Journal of Medical Toxicology
2011
, vol. 
7
 (pg. 
105
-
108
)
7
Vorce
S.P.
Holler
J.M.
Cawrse
B.M.
Magluilo
J.
Dimethylamylamine: a drug causing positive immunoassay results for amphetamines
Journal of Analytical Toxicology
2011
, vol. 
35
 (pg. 
183
-
187
)
8
Yee
L.M.
Wu
D.
False-positive amphetamine toxicology screen results in three pregnant women using labetalol
Obstetrics and Gynecology
2011
, vol. 
117
 
2 Pt 2
(pg. 
503
-
506
)
9
Fucci
N.
False positive results for amphetamine in urine of a patient with diabetes mellitus
Forensic Science International
2012
, vol. 
223
 pg. 
e60
 
10
Nomier
M.A.
AL-Huseini
H.K.
False-positive TDxFLx urine amphetamine/methamphetamine II assay from ofloxacin
Saudi Pharmaceutical Journal
2004
, vol. 
12
 (pg. 
42
-
46
)
11
Logan
B.K.
Costantino
A.G.
Rieders
E.F.
Sanders
D.
Trazodone, meta-chlorophenylpiperazine (an hallucinogenic drug and trazodone metabolite), and the hallucinogen trifluoromethylphenylpiperazine cross-react with the EMIT® II Ecstasy immunoassay in urine
Journal of Analytical Toxicology
2010
, vol. 
34
 (pg. 
587
-
589
)
12
Baron
J.M.
Griggs
D.A.
Nixon
A.L.
Long
W.H.
Flood
J.G.
The trazodone metabolite meta-chlorophenylpiperazine can cause false-positive urine amphetamine immunoassay results
Journal of Analytical Toxicology
2011
, vol. 
35
 (pg. 
364
-
368
)
13
Melanson
S.E.
Lewandrowski
E.L.
Griggs
D.A.
Flood
J.G.
Interpreting tricyclic antidepressant measurements in urine in an emergency department setting: comparison of two qualitative point-of-care urine tricyclic antidepressant drug immunoassays with quantitative serum chromatographic analysis
Journal of Analytical Toxicology
2007
, vol. 
31
 (pg. 
270
-
275
)
14
Matos
M.E.
Burns
M.M.
Shannon
M.W.
False-positive tricyclic antidepressant drug screen results leading to the diagnosis of carbamazepine intoxication
Pediatrics
2000
, vol. 
105
 pg. 
E66
 
15
Poklis
A.
Edinboro
L.E.
Lee
J.S.
Crooks
C.R.
Evaluation of a colloidal metal immunoassay device for the detection of tricyclic antidepressants in urine
Journal of Toxicology Clinical Toxicology
1997
, vol. 
35
 (pg. 
77
-
82
)
16
Sloan
K.L.
Haver
V.M.
Saxon
A.J.
Quetiapine and false-positive urine drug testing for tricyclic antidepressants
American Journal of Psychiatry
2000
, vol. 
157
 (pg. 
148
-
149
)
17
Hendrickson
R.G.
Morocco
A.P.
Quetiapine cross-reactivity among three tricyclic antidepressant immunoassays
Journal of Toxicology – Clinical Toxicology
2003
, vol. 
41
 (pg. 
105
-
108
)
18
Chathanchirayil
S.J.
False positive urine drug screening for tricyclic antidepressants in patients taking quetiapine
Australian and New Zealand Journal of Psychiatry
2011
, vol. 
45
 pg. 
792
 
19
Blank
A.
Hellstern
V.
Schuster
D.
Hartmann
M.
Matthee
A.K.
Burhenne
J.
, et al. 
Efavirenz treatment and false-positive results in benzodiazepine screening tests
Clinical Infectious Diseases: An Official Publication of the Infectious Diseases Society of America
2009
, vol. 
48
 (pg. 
1787
-
1789
)
20
Nasky
K.M.
Cowan
G.L.
Knittel
D.R.
False-positive urine screening for benzodiazepines: an association with sertraline? A two-year retrospective chart analysis
Psychiatry (Edgmont)
2009
, vol. 
6
 (pg. 
36
-
39
)
21
Mikel
C.
Pesce
A.J.
Rosenthal
M.
West
C.
Therapeutic monitoring of benzodiazepines in the management of pain: current limitations of point of care immunoassays suggest testing by mass spectrometry to assure accuracy and improve patient safety
Clinica Chimica Acta; International Journal of Clinical Chemistry
2012
, vol. 
413
 (pg. 
1199
-
1202
)
22
Rossi
S.
Yaksh
T.
Bentley
H.
van den Brande
G.
Grant
I.
Ellis
R.
Characterization of interference with 6 commercial Δ9-tetrahydrocannabinol immunoassays by efavirenz (glucuronide) in urine
Clinical Chemistry
2006
, vol. 
52
 (pg. 
896
-
897
)
23
Oosthuizen
N.M.
Laurens
J.B.
Efavirenz interference in urine screening immunoassays for tetrahydrocannabinol
Annals of Clinical Biochemistry
2012
, vol. 
49
 
Pt 2
(pg. 
194
-
196
)
24
Rollins
D.
Jennison
T.
Jones
G.
Investigation of interference by nonsteroidal anti-inflammatory drugs in urine tests for abused drugs
Clinical Chemistry
1990
, vol. 
36
 (pg. 
602
-
606
)
25
Kovatsi
L.
Pouliopoulos
A.
Papadaki
A.
Samanidou
V.
Tsoukali
H.
Development and validation of a high-performance liquid chromatography method for the evaluation of niflumic acid cross-reactivity of two commercial immunoassays for cannabinoids in urine
Journal of Analytical Toxicology
2010
, vol. 
34
 (pg. 
229
-
232
)
26
Cotten
S.W.
Duncan
D.L.
Burch
E.A.
Seashore
C.J.
Hammett-Stabler
C.A.
Unexpected interference of baby wash products with a cannabinoid (THC) immunoassay
Clinical Biochemistry
2012
, vol. 
45
 (pg. 
605
-
609
)
27
Janecka
A.
Fichna
J.
Janecki
T.
Opioid receptors and their ligands
Current Topics in Medicinal Chemistry
2004
, vol. 
4
 (pg. 
1
-
17
)
28
Straseski
J.A.
Stolbach
A.
Clarke
W.
Opiate-positive immunoassay screen in a pediatric patient
Clinical Chemistry
2010
, vol. 
56
 (pg. 
1220
-
1223
)
29
Birch
M.A.
Couchman
L.
Pietromartire
S.
Karna
T.
Paton
C.
McAllister
R.
, et al. 
False-positive buprenorphine by CEDIA in patients prescribed amisulpride or sulpiride
Journal of Analytical Toxicology
2013
, vol. 
37
 (pg. 
233
-
236
)
30
Pavlic
M.
Libiseller
K.
Grubwieser
P.
Rabl
W.
Cross-reactivity of the CEDIA buprenorphine assay with opiates: an Austrian phenomenon?
International Journal of Legal Medicine
2005
, vol. 
119
 (pg. 
378
-
381
)
31
Fucci
N.
EMIT dau: methadone false positive with creatinine interference
Forensic Science International
2005
, vol. 
148
 pg. 
81
 
32
Bottcher
M.
Beck
O.
Evaluation of buprenorphine CEDIA assay versus GC-MS and ELISA using urine samples from patients in substitution treatment
Journal of Analytical Toxicology
2005
, vol. 
29
 (pg. 
769
-
776
)
33
Rogers
S.C.
Pruitt
C.W.
Crouch
D.J.
Caravati
E.M.
Rapid urine drug screens: diphenhydramine and methadone cross-reactivity
Pediatric Emergency Care
2010
, vol. 
26
 (pg. 
665
-
666
)
34
Schneider
S.
Wennig
R.
Interference of diphenhydramine with the EMIT II immunoassay for propoxyphene
Journal of Analytical Toxicology
1999
, vol. 
23
 (pg. 
637
-
638
)
35
Baden
L.R.
Horowitz
G.
Jacoby
H.
Eliopoulos
G.M.
Quinolones and false-positive urine screening for opiates by immunoassay technology
JAMA: The Journal of the American Medical Association
2001
, vol. 
286
 (pg. 
3115
-
3119
)
36
Tenore
P.L.
False-positive buprenorphine EIA urine toxicology results due to high dose morphine: a case report
Journal of Addictive Diseases
2012
, vol. 
31
 (pg. 
329
-
331
)
37
Backmund
M.
Meyer
K.
Zielonka
M.
Eichenlaub
D.
Ofloxacin causes false-positive immunoassay results for urine opiates
Addiction Biology
2000
, vol. 
5
 (pg. 
319
-
320
)
38
Meatherall
R.
Dai
J.
False-positive EMIT II opiates from ofloxacin
Therapeutic Drug Monitoring
1997
, vol. 
19
 (pg. 
98
-
99
)
39
Jenkins
A.J.
Lavins
E.S.
Snyder
A.
Evaluation of the Cedia heroin metabolite (6-AM) immunoassay with urine specimens from A criminal justice drug-testing program
Journal of Analytical Toxicology
2005
, vol. 
29
 (pg. 
201
-
204
)
40
Lancelin
F.
Kraoul
L.
Flatischler
N.
Brovedani-Rousset
S.
Piketty
M.-L.
False-positive results in the detection of methadone in urines of patients treated with psychotropic substances
Clinical Chemistry
2005
, vol. 
51
 (pg. 
2176
-
2177
)
41
Widschwendter
C.G.
Zernig
G.
Hofer
A.
Quetiapine cross reactivity with urine methadone immunoassays
American Journal of Psychiatry
2007
, vol. 
164
 pg. 
172
 
42
Fischer
M.
Reif
A.
Polak
T.
Pfuhlmann
B.
Fallgatter
A.J.
False-positive methadone drug screens during quetiapine treatment
Journal of Clinical Psychiatry
2010
, vol. 
71
 pg. 
1696
 
43
Lasic
D.
Uglesic
B.
Zuljan-Cvitanovic
M.
Supe-Domic
D.
Uglesic
L.
False-positive methadone urine drug screen in a patient treated with quetiapine
Acta Clinica Croatica
2012
, vol. 
51
 (pg. 
269
-
272
)
44
de Paula
M.
Saiz
L.C.
Gonzalez-Revalderia
J.
Pascual
T.
Alberola
C.
Miravalles
E.
Rifampicin causes false-positive immunoassay results for urine opiates
Clinical Chemistry and Laboratory Medicine
1998
, vol. 
36
 (pg. 
241
-
243
)
45
Collins
A.A.
Merritt
A.P.
Bourland
J.A.
Cross-reactivity of tapentadol specimens with DRI methadone enzyme immunoassay
Journal of Analytical Toxicology
2012
, vol. 
36
 (pg. 
582
-
587
)
46
Shaikh
S.
Hull
M.J.
Bishop
K.A.
Griggs
D.A.
Long
W.H.
Nixon
A.L.
, et al. 
Effect of tramadol use on three point-of-care and one instrument-based immunoassays for urine buprenorphine
Journal of Analytical Toxicology
2008
, vol. 
32
 (pg. 
339
-
343
)
47
Lichtenwalner
M.R.
Mencken
T.
Tully
R.
Petosa
M.
False-positive immunochemical screen for methadone attributable to metabolites of verapamil
Clinical Chemistry
1998
, vol. 
44
 (pg. 
1039
-
1041
)
48
Kosten
T.R.
Fiellin
D.A.
U.S. National Buprenorphine Implementation Program
Buprenorphine for office-based practice: consensus conference overview
The American Journal on Addictions/American Academy of Psychiatrists in Alcoholism and Addictions
2004
, vol. 
13
 
Suppl 1
(pg. 
S1
-
S7
)
49
Cherwinski
K.
Petti
T.A.
Jekelis
A.
False methadone-positive urine drug screens in patients treated with quetiapine
Journal of the American Academy of Child & Adolescent Psychiatry
2007
, vol. 
46
 (pg. 
435
-
436
)
50
Geraci
M.J.
Peele
J.
McCoy
S.L.
Elias
B.
Phencyclidine false positive induced by lamotrigine (Lamictal®) on a rapid urine toxicology screen
International Journal of Emergency Medicine
2010
, vol. 
3
 (pg. 
327
-
331
)
51
Macher
A.M.
Penders
T.M.
False-positive phencyclidine immunoassay results caused by 3,4-methylenedioxypyrovalerone (MDPV)
Drug Testing and Analysis
2013
, vol. 
5
 (pg. 
130
-
132
)
52
Hull
M.J.
Griggs
D.
Knoepp
S.M.
Smogorzewska
A.
Nixon
A.
Flood
J.G.
Postmortem urine immunoassay showing false-positive phencyclidine reactivity in a case of fatal tramadol overdose
The American Journal of Forensic Medicine and Pathology
2006
, vol. 
27
 (pg. 
359
-
362
)
53
Ly
B.T.
Thornton
S.L.
Buono
C.
Stone
J.A.
Wu
A.H.
False-positive urine phencyclidine immunoassay screen result caused by interference by tramadol and its metabolites
Annals of Emergency Medicine
2012
, vol. 
59
 (pg. 
545
-
547
)
54
King
A.M.
Pugh
J.L.
Menke
N.B.
Krasowski
M.D.
Lynch
M.J.
Pizon
A.F.
Nonfatal tramadol overdose may cause false-positive phencyclidine on Emit-II assay
American Journal of Emergency Medicine
2013
, vol. 
31
 
444
(pg. 
e5
-
9
)
55
Sena
S.F.
Kazimi
S.
Wu
A.H.
False-positive phencyclidine immunoassay results caused by venlafaxine and O-desmethylvenlafaxine
Clinical Chemistry
2002
, vol. 
48
 (pg. 
676
-
677
)
56
Bond
G.R.
Steele
P.E.
Uges
D.R.
Massive venlafaxine overdose resulted in a false positive Abbott AxSYM urine immunoassay for phencyclidine
Journal of Toxicology – Clinical Toxicology
2003
, vol. 
41
 (pg. 
999
-
1002
)
57
Santos
P.M.
Lopez-Garcia
P.
Navarro
J.S.
Fernandez
A.S.
Sadaba
B.
Vidal
J.P.
False positive phencyclidine results caused by venlafaxine
American Journal of Psychiatry
2007
, vol. 
164
 pg. 
349
 
58
Rohrich
J.
Zorntlein
S.
Lotz
J.
Becker
J.
Kern
T.
Rittner
C.
False-positive LSD testing in urine samples from intensive care patients
Journal of Analytical Toxicology
1998
, vol. 
22
 (pg. 
393
-
395
)
59
Ritter
D.
Cortese
C.M.
Edwards
L.C.
Barr
J.L.
Chung
H.D.
Long
C.
Interference with testing for lysergic acid diethylamide
Clinical Chemistry
1997
, vol. 
43
 (pg. 
635
-
637
)
60
Wiegand
R.F.
Klette
K.L.
Stout
P.R.
Gehlhausen
J.M.
Comparison of EMIT® II, CEDIA® and DPC® RIA assays for the detection of lysergic acid diethylamide in forensic urine samples
Journal of Analytical Toxicology
2002
, vol. 
26
 (pg. 
519
-
523
)
61
Gagajewski
A.
Davis
G.K.
Kloss
J.
Poch
G.K.
Anderson
C.J.
Apple
F.S.
False-positive lysergic acid diethylamide immunoassay screen associated with fentanyl medication
Clinical Chemistry
2002
, vol. 
48
 (pg. 
205
-
206
)
62
Citterio-Quentin
A.
Seidel
E.
Ramuz
L.
Parant
F.
Moulsma
M.
LSD screening in urine performed by CEDIA LSD assay: positive interference with sertraline
Journal of Analytical Toxicology
2012
, vol. 
36
 (pg. 
289
-
290
)